|


| |
The following IBASIC software control the
built-in functions of the analyzer as well as provides additional capabilities to the
analyzer. The analyzer must be equipped with the Instrument BASIC option and firmware
revision A.00.04 or higher. The software is compatible with the two and four input channel
configurations.
HCAL
(Hand-Help Calibration)
 |
HCAL Background
Hand-held calibration of accelerometers is a very simple and
straightforward method. A primary advantage of hand-held calibration is that it can
perform an end-to-end calibration of the transducer and input channel of the data
acquisition system. A hand-held calibrator is a small, portable, battery- powered device.
The system consists of an electromagnetic exciter driven by a oscillator at a constant
frequency, (typically 79.6 or 159.0 Hertz, nominal). An accelerometer is mounted on top of
the calibrator and a built-in reference accelerometer controls a servo-feedback loop to
maintain a constant reference level (typically 1 g rms or 10 m/s2
rms, nominal) for calibration. Please click on Thumbnail
to enlarge. |
RCAL
(Ratio Calibration)
 |
RCAL Background
Static calibration of the force sensor of an impactor is not satisfactory
because the sensitivity is dependent on several factors, including the type of tip and the
mass of the impactor. The recommended method of calibrating transducers for impact testing
is ratio calibration [Halvorsen and Brown, Sound and Vibration, Nov. 1977]. With
this method, the frequency response of a large inertial mass is measured by mounting an
accelerometer on one side of it and impacting on the other side. The calibration FRF can
be used as a calibration curve or a single, averaged value can be determined for a given
frequency range as a calibration value.
When measuring frequency response functions
with impact testing, the absolute calibration values of the transducers are not necessary,
but it is convenient to calibrate the ratio of the force sensor and accelerometer. In this
case, the calibration value of the force sensor is typically set to unity, and a number of
accelerometers are calibrated relative to the impactor. To determine the absolute
calibration of the impactor, an accelerometer that has been calibrated by some other
method serves as the reference transducer. Please click
on Thumbnail to enlarge. |
DCAL
(Drop Calibration)
 |
DCAL Background
Drop calibration, which is also known as free-fall calibration, is a very
basic and natural method of calibrating accelerometers and force sensors. It is based on a
predictable and repeatable motion, that a body in free-fall experiences a constant
gravitational acceleration, and by measuring the step response of the transducer to the
one-g input, its sensitivity is determined. An actual drop calibration response
differs from the ideal free-fall step response in several aspects, and the measured data
is curve-fit to determine the sensitivity. Drop calibration is an absolute calibration
method and does not require a reference standard because the results are referenced to the
local gravitational constant. Please click on Thumbnail
to enlarge. |
CHANREF (Channel Reference) program
| |
CHANREF puts point number and directions into the header of the SDF files.
|
RRIT
(Roving Response Impact
Testing)
 |
RRIT Background
Roving Response Impact Testing
is a modal data
acquisition method in which frequency response functions are measured between a fixed
impactor and one or more roving response transducers. The basic procedure is that the test
system is repeatedly impacted at an input position while the response transducers are
roved about the measurement positions. RRIT is a single reference technique, but to
collect multiple reference data with RRIT, the procedure is repeated for other input
positions. Alternatively, the impacts could be made at all input positions before
relocating the response transducers. Since only one input is applied at a time, a
multiple-input FRF estimator is not required. The RRIT method is a convenient and
practical technique that is very well suited for troubleshooting vibration problems and
field testing. Additional information on impact testing procedures are included in the
relevant sections of this manual. Please click on Thumbnail
to enlarge. |
MRIT (Multiple Reference Impact Testing)
 |
MRIT Background
Multiple Reference Impact Testing
(MRIT) does not imply impacting a
test structure with more than one hammer. Instead, the outputs are taken as the
references, rather than the input. MRIT is a modal data acquisition method in which
frequency response functions (FRF) are measured between a single, roving impactor and
multiple, fixed response transducers. Possible reference transducers are accelerometers,
microphones, and proximity probes. A set of multiple reference FRFs is generated by
applying the reciprocity relationships to determine the response at the impact locations.
The set of measurements can then be treated as multiple input data and processed with
multiple reference modal parameter estimation software. Please
click on Thumbnail to enlarge. |
Agilent is a registered trademark of Agilent
Technologies.
For Application Questions and
United States and Canada Sales,
please contact: |
Navcon Engineering Network
Telephone: (+1) 714-441-3488
FAX: (+1) 714-441-3487 |
Questions / Comments - please send mail to
webinfo@navcon.com ..
Copyright © 1997 - 2007
Navcon Engineering Network
Last modified:
March 19, 2013
|